| Model Number |
MMT-522NAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that insulin pump alarmed compromised force sensor system and squirting out during manual prime.Drive support cap was sticking out.Customer¿s blood glucose was 150mg/dl at time of incident.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The device received compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.No moisture damage on electronics and motor assembly noted.
|
| |
|