MEDTRONIC MINIMED PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522NAL
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call that insulin pump alarmed compromised force sensor system and squirting out during manual prime.Drive support cap was sticking out.Customer¿s blood glucose was 150mg/dl at time of incident.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
The device received compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.No moisture damage on electronics and motor assembly noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key7341290
Report Number2032227-2018-01951
Device Sequence Number16339874
Product Code OYC
UDI-Device Identifier00613994223685
UDI-Public(01)00613994223685(017)20091231
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 09/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/31/2009
Device Model NumberMMT-522NAL
Device Catalogue NumberMMT-522NAL
Device Lot NumberA000093859
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2018
Initial Date Received by Manufacturer 03/05/2018
Supplement Date Received by Manufacturer08/15/2018
Initial Report FDA Received Date03/15/2018
Supplement Report FDA Received Date09/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age47 YR
Patient Weight195
-
-