STRYKER ENDOSCOPY-SAN JOSE PKG., TOMCAT SHAVER BLADE, F - SERIES - 4.0MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Catalog Number 0375545000 |
Device Problem
Device Packaging Compromised (2916)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/21/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.(b)(4).
|
|
Event Description
|
It was reported that foreign material was found inside in sterile packaging.No delay and no adverse consequences were reported.
|
|
Manufacturer Narrative
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: when opening the pouch, the nurse observed the presence of a little green plastic in the bur shaft.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be an operator accidentally scraped the sharp part of the cutter with the gloves during assembly and remained attached to the unit, improper cleaning process, qc incoming and in-process inspections.Gtin: (b)(4).
|
|
Event Description
|
It was reported that foregin material was found inside in sterile packaging.No delay and no adverse consequences were reported.
|
|