Catalog Number 0375545000
Device Problem Device Packaging Compromised (2916)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/21/2018
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.(b)(4).
 
Event Description
It was reported that foreign material was found inside in sterile packaging.No delay and no adverse consequences were reported.
 
Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: when opening the pouch, the nurse observed the presence of a little green plastic in the bur shaft.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be an operator accidentally scraped the sharp part of the cutter with the gloves during assembly and remained attached to the unit, improper cleaning process, qc incoming and in-process inspections.Gtin: (b)(4).
 
Event Description
It was reported that foregin material was found inside in sterile packaging.No delay and no adverse consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., TOMCAT SHAVER BLADE, F - SERIES - 4.0MM
Type of Device
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
valerie estrada
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7342442
MDR Text Key102732336
Report Number0002936485-2018-00254
Device Sequence Number1
Product Code GFA
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0375545000
Device Lot Number13023CE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-