MAUDE Adverse Event Report: STRYKER SPINE-US XIA 3 TITANIUM ILIAC SCREWDRIVER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US XIA 3 TITANIUM ILIAC SCREWDRIVER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 48231326
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/08/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that too much torque was applied to closed head iliac screwdrivers.There are no reports of adverse health consequences.
 
Additional Manufacturer Narrative
Method: visual inspection, device history review, complaint history review, risk assessment.Result: the customer reported event was confirmed via visual inspection.The outer sleeve was returned and it was found that the threading was fractured.The fractured pieces were not returned.Manufacturing history was reviewed and no issues were identified.Material analysis on similar devices found the tip fractured via ductile overload under an applied torsional force.Conclusion: according to similar material analysis and the event description, the most likely root cause is excessive force was applied to the driver causing it to fracture and deform.It is possible improper screw hole preparation, hard bone quality, under tapping, or misalignment of the driver to the screw contributed to the event as well.
 
Event or Problem Description
It was reported that too much torque was applied to closed head iliac screwdrivers.There are no reports of adverse health consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 TITANIUM ILIAC SCREWDRIVER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7342529
Report Number0009617544-2018-00068
Device Sequence Number13592018
Product Code LXH
UDI-Device Identifier07613327265279
UDI-Public(01)07613327265279
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number48231326
Device Catalogue Number48231326
Device Lot Number140957
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Initial Date Received by Manufacturer 02/14/2018
Supplement Date Received by Manufacturer04/26/2018
Initial Report FDA Received Date03/15/2018
Supplement Report FDA Received Date05/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-