Method: visual inspection, device history review, complaint history review, risk assessment.Result: the customer reported event was confirmed via visual inspection.The inner shaft was returned and it was confirmed that the tip was deformed.The threading was twisted and misshapen.Manufacturing history was reviewed and no issues were identified.Material analysis on similar devices found the tip fractured via ductile overload under an applied torsional force.Conclusion: according to similar material analysis and the event description, the most likely root cause is excessive force was applied to the driver causing it to fracture and deform.It is possible improper screw hole preparation, hard bone quality, under tapping, or misalignment of the driver to the screw contributed to the event.
|