• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 32/0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 32/0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 6570-0-132
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
It was reported that in performing a primary hip (side not reported) procedure, the 32mm +0 head would not lock onto the accolade ii stem.The head was able to be manually removed by the surgeon.Another head was immediately available and locked onto the stem with no issues.Rep reported that no case time was lost (no surgical delay).
 
Additional Manufacturer Narrative
An event regarding a size/fit issue involving a ceramic head was reported.The event was not confirmed.Dimensional inspection: a dimensional inspection was performed on the returned device where all dimensions were found to be within specification.Functional inspection: not performed as the exact failure mode could not be replicated.Material analysis: not performed as the reported event is not related to material integrity of the device.Medical records received and evaluation: a medical review was not performed because no medical information was provided.Device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: review indicated there have been no other similar events for the reported lot.Conclusions: the event could not be confirmed nor the root cause determined because insufficient information was provided.Further information such as operative reports are needed to investigate this event further.If additional information becomes available, this investigation will be reopened.
 
Event or Problem Description
It was reported that in performing a primary hip (side not reported) procedure, the 32mm +0 head would not lock onto the accolade ii stem.The head was able to be manually removed by the surgeon.Another head was immediately available and locked onto the stem with no issues.Rep reported that no case time was lost (no surgical delay).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA V-40 CERAMIC HEAD 32/0
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7342834
Report Number0002249697-2018-00764
Device Sequence Number1657746
Product Code LZO
UDI-Device Identifier04546540608505
UDI-Public04546540608505
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K052718
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Catalogue Number6570-0-132
Device Lot Number62189001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2018
Initial Date Received by Manufacturer 02/19/2018
Supplement Date Received by Manufacturer07/09/2018
Initial Report FDA Received Date03/15/2018
Supplement Report FDA Received Date08/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-