The explanted lead was received for analysis.Product analysis on the lead was completed and approved.An analysis was performed on the returned lead portion and the reported allegations of lead fracture were confirmed.Note that the majority of the lead assembly (body) including the electrodes was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis the end of the connector pin quadfilar coil appeared to be broken approximately 122 mm from the end of the connector boot.Scanning electron microscopy was performed and identified the area as having evidence of being worn to the point of fracture with flat spots and pitting on the coil surface.It is unknown if the break occurred while stimulation was present due to the absence of metal pitting on the broken coil wire surfaces.The abraded openings found on the outer and inner silicone tubes, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer and inner silicone tubes.With the exception of the discontinuity and abraded openings, the condition of the returned lead portion is consistent with those that typically exist following an explant procedure.No other obvious anomalies were noted.Continuity checks of the returned lead portion were performed, during the visual analysis, and no discontinuities were identified.No additional or relevant information has been received to date.
|