| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a temperature issue.The reporter alleged the pump was "hot," and no damage was found to the pump.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because there is the potential for the user to experience an injury to the skin due to increased pump temperature.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up # 1 date of submission 20-apr-2018.Device evaluation: the device has been returned and evaluated by product analysis on 13-apr-2018 with the following findings: a review of the pump¿s black box and download history did not find any activity related to the complaint.The battery compartment and battery cap were intact and the battery cap could be fully tightened onto the pump.The pump powered on normally with no overheating or alarms duplicated during the investigation.The pump¿s ¿ez-prime¿ steps were performed correctly and all of the pump's electrical current draws were found to be within specification.The pump cover was removed and there were no intermittent condition found to the power circuit.The investigation was unable to duplicate the initial ¿temperature¿ complaint.
|
| |
|