JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a pcb00 13.0 diopter intraocular lens (iol) was implanted in the patient's operative eye on (b)(6) 2017.It was later explanted on (b)(6) 2017 because the lens made the patient too near sighted.There was no incision enlargement, but a suture was used.The replacement lens used was a different model.Reportedly, there was no patient injury.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 03/22/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual evaluation reveals the lens returned is a cut three piece lens (a pcb00 was reported as the product associated to the complaint); therefore, the product associated to the complaint does not match the reported issue.The reported issue was not verified.The complaint product was not returned.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no additional complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7343173
Report Number2648035-2018-00383
Device Sequence Number173554
Product Code HQL
UDI-Device Identifier05050474558120
UDI-Public(01)05050474558120(17)191019
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date10/19/2019
Device Model NumberPCB00
Device Catalogue NumberPCB0000130
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2018
Initial Date Received by Manufacturer 02/15/2018
Supplement Date Received by Manufacturer04/18/2018
Initial Report FDA Received Date03/15/2018
Supplement Report FDA Received Date04/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-