Model Number PCB00
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Visual Impairment (2138); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a pcb00 9.5 diopter intraocular lens (iol) was implanted in the patient's operative eye.It was later explanted due to decreased visual acuity and photophobia.The replacement lens was the same model, but different diopter of 18.5.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes, returned to manufacturer on: 03/22/2018, device returned to manufacturer? yes.Device evaluation: the sample was received in specimen container.Only the lens from the preload delivery system was received for evaluation.Visual inspection using microscope magnification was performed: the optic section of the lens was cut in three (3) portions.Both haptics were observed detached.Viscoelastic residue was observed in the lens pieces.The condition observed in the lens and the way in which the cut was formed is consistent with a unit that has been cut due to explant process.The returned lens parts were observed clear as expected in the 1-piece acrylic lens.Only viscoelastic was detected.The reported issue could not be verified on the returned lens.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.There was no discrepancy found during the review.A search revealed that no other complaint was received from this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7343376
Report Number2648035-2018-00385
Device Sequence Number147394
Product Code HQL
UDI-Device Identifier05050474558052
UDI-Public(01)05050474558052(17)200531
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date05/31/2020
Device Model NumberPCB00
Device Catalogue NumberPCB0000095
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2018
Initial Date Received by Manufacturer 02/15/2018
Supplement Date Received by Manufacturer04/16/2018
Initial Report FDA Received Date03/15/2018
Supplement Report FDA Received Date05/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
-
-