| Model Number |
MMT-523RNAB |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had an unexpected restart.The customer's blood glucose level was 365 mg/dl.The customer was assisted with the troubleshooting.The insulin pump will be returned for the analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Unit power up properly after battery installation.No blank display noted.Unit was received with normal operating currents and passed self-test, unexpected restart error test and displacement test.No unexpected restart, off no power, low battery, weak battery, battery out limit and failed battery test alarms noted during testing.
|
| |
|