MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751RNAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751RNAS
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 02/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they experienced high blood glucose level of 460 mg/dl and also no delivery alarm on insulin pump.The customer¿s current blood glucose was 431 mg/dl.The customer treated high blood glucose with manual injections.The product was returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump passed displacement test, rewind test, basic occlusion test, prime test, excessive no delivery test and occlusion test.No excessive no delivery alarms noted.Insulin pump meets specification: yes the information provided in concomitant reporting section with the initial report was incorrect.The correct information has been included with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751RNAS
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7343951
Report Number3004209178-2018-57261
Device Sequence Number679615
Product Code OZO
UDI-Device Identifier00643169508002
UDI-Public(01)00643169508002
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 12/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-751RNAS
Device Catalogue NumberMMT-751RNAS
Device Lot NumberA3751RNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2018
Initial Date Received by Manufacturer 02/26/2018
Supplement Date Received by Manufacturer12/21/2018
Initial Report FDA Received Date03/15/2018
Supplement Report FDA Received Date12/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age52 YR
Patient Weight190
-
-