| Model Number |
MMT-751RNAS |
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
02/24/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced high blood glucose level of 460 mg/dl and also no delivery alarm on insulin pump.The customer¿s current blood glucose was 431 mg/dl.The customer treated high blood glucose with manual injections.The product was returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed displacement test, rewind test, basic occlusion test, prime test, excessive no delivery test and occlusion test.No excessive no delivery alarms noted.Insulin pump meets specification: yes the information provided in concomitant reporting section with the initial report was incorrect.The correct information has been included with this report.
|
| |
|