Model Number ZCB00
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens was not implanted.(b)(6).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that immediately after taking out the intraocular lens (iol) from the lens case, the lens started cracking through the whole diameter of the optic.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation, returned to manufacturer on: 04/05/2018.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection under the microscope showed loose fibers/particles and residue of lubricant material were observed on the lens pieces related to the handling of the unit out of a sterile environment.The lens was observed with two cuts on the lens edge section.A small depression mark was also observed on the lens surface.No damage was observed on the haptics.The condition of the returned lens was consistent with a product that was handled and prepared for surgical process.The customer's reported complaint was verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional investigation requests for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation, there is no indication of a product quality deficiency and the reported issue was verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7344221
Report Number2648035-2018-00390
Device Sequence Number16160548
Product Code HQL
UDI-Device Identifier05050474531642
UDI-Public(01)05050474531642(17)210504
Combination Product (Y/N)N
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date05/04/2021
Device Model NumberZCB00
Device Catalogue NumberZCB0000165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2018
Initial Date Received by Manufacturer 02/16/2018
Supplement Date Received by Manufacturer04/09/2018
Initial Report FDA Received Date03/16/2018
Supplement Report FDA Received Date05/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-