Model Number 5300020901
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow up report will be filed once the quality investigation is complete.
 
Event or Problem Description
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
 
Additional Manufacturer Narrative
The reported bur involved with this event was returned for evaluation and the reported failure of bur breakage was confirmed.The device was sent to the product engineering who concluded that the bur failed as a result of excessive bending or impact force.
 
Event or Problem Description
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.0MM OVAL SOLID CARBIDE BUR LONG
Common Device Name
BUR, SURGICAL, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key7344811
Report Number0001811755-2018-00522
Device Sequence Number1054819
Product Code GFF
UDI-Device Identifier07613153001126
UDI-Public(01)07613153001126
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/01/2022
Device Model Number5300020901
Device Catalogue Number5300020901
Device Lot Number17250017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2018
Initial Date Received by Manufacturer 02/20/2018
Supplement Date Received by Manufacturer05/23/2018
Initial Report FDA Received Date03/16/2018
Supplement Report FDA Received Date05/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates