| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned to animas for evaluation.An evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.The cartridge compartment cap was alleged to be cracked/damaged.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the cartridge compartment.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission (b)(6) 2018 device evaluation: the device has been returned and evaluated by product analysis on (b)(6) 2018 with the following findings: during investigation, the cartridge compartment was found to be cracked.Unrelated to the original complaint, the battery compartment was found to be cracked starting at the threads to the left side of the grip pad down to the seal and from the case seal traveling to the grip pad.The threads of the returned battery cap were found to be stripped.The returned battery cap was unable to properly secure to the returned pump or a test pump.A test battery cap was used and able to properly secure to the pump.
|
| |
|