Catalog Number 8880928
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent an unspecified surgery.Intra-op, the implant broke by sheer force when it was being implanted.The peek portion caught and was steered from the ti portion.The implant was replaced by the same size implant and was implanted without issue.All the fragments of the broken implant were removed.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis: visual, microscopic and functional evaluation was performed.The elevate spacer has broken from what appears to be material overload.There are multiple witness marks on the implant that are consistent with impaction.In its as returned state, the nose of the implant would not have been fully closed.It is not known if the implant was in this position when there was an attempt made to install the spacer.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATE SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7345084
Report Number1030489-2018-00409
Device Sequence Number13601174
Product Code MAX
UDI-Device Identifier00643169430822
UDI-Public00643169430822
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K142559
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/18/2024
Device Catalogue Number8880928
Device Lot Number0508004W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2018
Initial Date Received by Manufacturer 02/22/2018
Supplement Date Received by Manufacturer04/25/2018
Initial Report FDA Received Date03/16/2018
Supplement Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-