| Catalog Number |
8300.25G02 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/03/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device history lot review: there was nothing relative to the complaint.This lot met the specifications outlined for release.Sample analysis: no sample was returned.Conclusion: there is no evidence of any manufacturing issues that would result in a tip detachment.This is likely an isolated incident.We are unable to confirm since no sample was received.(b)(4).All available information has been provided.This complaint has now been closed.
|
| |
|
Event or Problem Description
|
|
A report has been received of an incident that occurred in france.During surgery, the tip detached from the handle before the end of the surgery.And during the surgery, there was a time with no light at all.There is no available information on the status of the patient therefore we assume no patient injury.
|
| |
|