| Model Number |
E SERIES |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device powered on without display.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the malfunction was duplicated and a system interconnection flex cable was replaced to remedy the problem.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|