| Model Number |
VLOCN0644 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic procedure, just in the beginning of the closure in peterson's hernia, the seal of the welded loop came loose, so the device could not be anchored.They used a new device to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned samples noted one suture and one needle.The product was received with the loop open.It was observed, under magnification, that the loop appeared to have split under tension.Upon microscopic inspection, evidence of material transfer was found at the weld site of the loop.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|