| Catalog Number |
309644 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that 10 units of the customers bd luer-lok¿ syringe with bd precisionglide¿ needle were cracked and leak.It is unspecified when the event occurred.No reports of serious injury or medical intervention noted.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: one loose 10ml luer lok syringe with no packaging was received by bd (b)(4) for investigation.Customer reported lot number #7300578, catalog # 309644.Based on visual examination of the returned sample, cracked barrel was observed and the crack was almost along the entire syringe barrel.Product defect confirmed based on sample evaluation performed.Dhr review for batch 7300578 (p/n 309644): manufacturing dates: 11/02/2017 to 11/03/2017.Batch quantity was (b)(4).Printing and assembly records were reviewed as part of this dhr review.All visual inspections were performed as per requirement with no quality notifications related to the complaint defect.Batch 7300578 was inspected and accepted based on meeting our inspection control plan and subsequently approved for shipment.The review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Root cause not required based on severity and occurrence level defined for this complaint.Based on the severity and occurrence a capa is not necessary.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Search Alerts/Recalls
|