• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER RADICAL-7 HANDHELD; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER RADICAL-7 HANDHELD; OXIMETER Back to Search Results
Model Number 21744
Device Problems Failure To Run On AC/DC (1001); Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/20/2018
Event Type  malfunction  
Manufacturer Narrative
The returned device was evaluated.The radical-7 handheld device passed visual inspection; however visual inspection of the docking station (rds) indicated a broken latch.The radical-7 functioned as designed in a known good docking station.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.The handheld was unable to stay docked due to the rds damage and as a result was unable to maintain communication and power between rds and radical-7.Due to loss of power the radical-7 will power off when battery power is depleted.
 
Event Description
The customer reported the device powers off and has as damaged housing.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADICAL-7 HANDHELD
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
Manufacturer (Section G)
INDUSTRIAL VALLERA DE MEXICALI S.A DE C.V.
calzada del oro, no. 2001
parque industrial palaco
mexicali, baja california 21600
MX   21600
Manufacturer Contact
erica kline
52 discovery
irvine, CA 92618-1604
9492977000
MDR Report Key7347462
MDR Text Key102906008
Report Number2031172-2018-00145
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140188
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21744
Device Catalogue Number9500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received02/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/07/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-