On (b)(6) 2016, a field service engineer (fse) visited the client site and replaced the defective all in one computer with its reporting software.He also loaded the printer driver and the keyspan usb serial driver.He tested the unit and the data transfer capability from the g8 to the reporting software and then to the laboratory information system (lis).The fse set up the computer according to work instructions 12-css-042-0 rev 13.In addition, he removed the hard drive from the defective computer and gave it to customer so that the client disposes of it according to their protocols.No further action was required by field service.The most probable cause of the reported event was defective all in one computer.(b)(4).This report is being submitted due to a retrospective review conducted under capa (b)(4).
|
On (b)(6) 2016, a customer reported a computer failure with their hlc-723g8 analyzer.The computer was in a repair-and-repeat startup repair cycle repeatedly.The customer was unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|