COOK INC TURBO-JECT SINGLE LUMEN PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
The international customer reported that the turbo-ject single lumen peripherally inserted central venous catheter leaked at the junction of the purple hub and the clear tubing on the distal end of the device.The picc line was utilized at least weekly for infusions of blood products and chemotherapy, as well as blood draws.The product problem occurred approximately 180 days after initial placement and necessitated the use of additional anesthetic and the placement of a new device in an additional procedure.The complaint device is reportedly available for return; however, as of the date of this report, no device has yet been received for evaluation.
 
Event or Problem Description
No additional information or changes at this time.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, device history record, documentation, instructions for use (ifu), trends, quality control, and visual inspection of the returned device was conducted during the investigation.The device was returned and the reported issue confirmed.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there was one other reported complaint for this lot number.Based on the information provided and the results of our investigation, a definitive root cause cannot be determined at this time, however appropriate measures have been initiated to address this failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-JECT SINGLE LUMEN PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET
Common Device Name
LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7348464
Report Number1820334-2018-00771
Device Sequence Number1556384
Product Code LJS
UDI-Device Identifier00827002558830
UDI-Public(01)00827002558830(17)190120(10)7606720
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K111244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberUPICS-3.0-CT-NT-1110
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/22/2018
Supplement Date Received by Manufacturer05/21/2018
Initial Report FDA Received Date03/19/2018
Supplement Report FDA Received Date06/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age5 YR
Patient Weight25
-
-