COOK INC STRAIGHT FIXED CORE WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Catalog Number TSF-25-145
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product code = dqx.Per the initial reporter, the device will not be returned.Pma/510(k) number = preamendment.(b)(4).This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that prior to use, three straight fixed core wire guides were observed to be incorrectly packaged.Per the reporter, normally the soft tip of the wire reaches out of the plastic holder when the product is opened; however, the stiff end of the wire was at the open end of the holder.The devices did not come into contact with a patient.It was noted that the devices could be utilized; however, incorrect packaging presented the potential for the wire guides to be misused.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT FIXED CORE WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7348465
Report Number1820334-2018-00641
Device Sequence Number10958850
Product Code DQX
UDI-Device Identifier00827002006171
UDI-Public(01)00827002006171(17)220213(10)7612750
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberTSF-25-145
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/07/2018
Supplement Date Received by Manufacturer05/22/2018
Initial Report FDA Received Date03/19/2018
Supplement Report FDA Received Date06/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-