• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC GUNTHER TULIP JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC GUNTHER TULIP JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-JUG
Device Problems Difficult to Remove (1528); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Edema (1820); Pain (1994); Tachycardia (2095); Perforation of Vessels (2135); Anxiety (2328); Distress (2329); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
It is alleged that the plaintiff received a gunther filter on (b)(6) 2005.It is alleged that the plaintiff was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
(b)(4).Additional information: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'tulip, venous insufficiency, tachycardia, edema (occlusion), dizziness, anxiety '.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Ivc thrombotic occlusion as an outcome for cook ivc filters is addressed in the published scientific literature.Ivc thrombotic occlusion is defined as the presence of an occluding thrombus in the ivc after filter insertion and documented by ultrasound (us), ct, mr imaging or venography; this may be symptomatic or asymptomatic.Unknown if the reported tachycardia, dizziness, and anxiety are directly related to the filter and unable to identify corresponding failure mode(s) at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
 
Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
This additional information received on 04apr2018 as follows: pt allegedly received an implant on (b)(6) 2005 due to gastric bypass procedure.Pt alleges venous insufficiency, tachycardia due to poor venous return secondary to ivc filter with peripheral edema and lower extremity edema.Pt further alleges dizziness and anxiety.
 
Manufacturer Narrative
The following fields were updated per additional information received.Appropriate term/code not available (3191) was selected for the alleged device perforation.Appropriate term/code not available (3191) was selected for the alleged device tilt investigation: the following allegations have been investigated: tilt, vena cava perforation, irretrievable filter, pain, loss of enjoyment of life, suffering, distress, limited mobility, leg swelling.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Physician practice guidelines and published guidance from regulatory agencies recommend that patients with indwelling filters undergo routine follow-up.The risks/benefits of filter retrieval should be considered for each patient during follow-up.Once protection from pe is no longer necessary, filter retrieval should be considered.Filter retrieval should be attempted when feasible and clinically indicated.Filter retrieval is a patient-specific, clinically complex decision; the decision to remove a filter should be based on each patient¿s individual risk/benefit profile (e.G., a patient¿s continued need for protection from pe compared to their experience with and (or) ongoing risk of experiencing filter-related complications).For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.The filter is designed to be retrieved with the günther tulip vena cava filter retrieval set.It may also be retrieved with the cloversnare® vascular retriever.Cook has not performed testing to evaluate the safety or effectiveness of filter retrieval using other retrieval systems or techniques.The published clinical literature includes descriptions of alternative techniques for filter retrieval; use of these techniques varies according to physician experience, patient anatomy, and filter position.The safety or effectiveness of these alternative retrieval techniques has not been established.Specific for ¿embedded¿ a filter that is embedded in the wall of the ivc may be difficult to retrieve.For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported pain, loss of enjoyment of life, suffering, distress, limited mobility, leg swelling are directly related to the filter and unable to identify a corresponding failure mode at this point in time.10 devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Patient is alleging tilt, vena cava perforation, irretrievable filter.Patient notes and further alleges "pain, suffering, loss of enjoyment of life, distress", limited activities "because if on the leg for long then it swells and is very painful, long travels can cause edema in lower extremities, tachycardia limits most physical activity, and cannot walk more than 20 minutes".08apr2019, per a report from computed tomography (ct); ¿findings: ivc filter: an intact infrarenal retrievable ivc filter is in place.One strut perforation measuring 4mm is noted to the right laterally without encroachment into bowel or other structure.The filter is moderately tilted to the left, however, the retrieval hook is not against the ivc wall.Impression: 1.Retrievable infrarenal ivc filter with moderate tilt and one strut perforation.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7348591
MDR Text Key102737893
Report Number1820334-2018-00754
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K073374
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup,Followup
Report Date 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-JUG
Was Device Available for Evaluation? No
Date Manufacturer Received02/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
Patient Weight64
-
-