• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE; INSULIN SYRINGE WITH NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE; INSULIN SYRINGE WITH NEEDLE Back to Search Results
Catalog Number 328411
Device Problem Unintended Movement (3026)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2018
Event Type  malfunction  
Manufacturer Narrative
The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd insulin syringe with the bd ultra-fine¿ needle plunger rod moves out of barrel on its own.Found during use.No reports of serious injury or medical intervention noted.
 
Manufacturer Narrative
Severity: s1; occurrence: a complaint history check was performed and this is the 1st related complaint reported for the defect/condition on lot number 7072878.Investigation summary: customer returned (1) loose 1cc, 12.7mm syringe.Customer states that that plunger rod moves out of barrel on its own.The returned syringe was examined and no defects were observed on the sample.The sample was tested and was able to draw and expel properly without any observed defects.A review of the device history record was completed for batch# 7072878.All inspections were performed per the applicable operations qc specifications.There was one (1) notification [200689255] noted that did not pertain to the complaint.Bd was not able to duplicate or confirm the customer¿s indicated failure.Root cause cannot be determined at this time as the issue is unconfirmed.Based on the above, no additional investigation and no capa is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE
Type of Device
INSULIN SYRINGE WITH NEEDLE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key7348634
MDR Text Key102864266
Report Number1920898-2018-00157
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
PMA/PMN Number
K170386
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number328411
Device Lot Number7072878
Initial Date Manufacturer Received 03/07/2018
Initial Date FDA Received03/19/2018
Supplement Dates Manufacturer Received03/07/2018
Supplement Dates FDA Received04/03/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-