| Catalog Number |
0064327 |
| Medical Device Problem Codes |
Computer Software Problem (1112); Patient Data Problem (3197)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An immucor specialist employee assessed the software in question on (b)(6) 2018 and confirmed that the date of birth demographic information is indeed carried over from the previous sample, but the immucor specialist employee was able to establish that no other demographic information is carried over.To clarify, this issue happens when a sample that does not have patient demographics is scanned into the qiacube worklist in immulink software following a sample that does have patient demographics.(b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a customer site reported a label generated from immulink software that contained the wrong patient demographic information.
|
| |
|