SMITH & NEPHEW, INC. ADOLESCENT TAN 8.5X40L; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
71645140 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/14/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a trauma surgery while placing the final nail after the second impact, the proximal tabs of the nail were opened and broken, the nail was inside the patient.With manual movements and a lot of caution, the doctor removed the nail from the patient.Procedure was completed with a new implant and no injury was reported.Delay was above 30 minutes due to the removal of the nail and the new implantation.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected legion ps xlpe high flexion insert was returned and evaluated.Our investigation included a lab analysis which indicated the nail fracture occurred proximally.The bone cement was applied to the entire length of the nail except the proximal end.The cement thickness was widely varied from proximal to the distal end.The fractured proximal keys were not returned.Based on the analysis conducted, the exact cause of the reported failure was not able to be determined.A clinical evaluation noted there is no clinical relevant documentation provided to perform a thorough medical investigation.No further medical assessment is warranted at this time.A review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the part revealed no prior complaints for the listed batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
|
| |
|
Search Alerts/Recalls
|
|
|