• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. SLEEK RX PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. SLEEK RX PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number 425-3022X
Medical Device Problem Codes Break (1069); Difficult to Insert (1316); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The investigation is currently underway.
 
Event or Problem Description
It was reported that the shaft of the pta balloon catheter allegedly broke during insertion through the sheath.The device was removed without incident.There was no reported difficulty in retracting the catheter through the introducer sheath.There was no reported patient injury.
 
Additional Manufacturer Narrative
The lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.The result of the investigation was confirmed.The evaluation of the returned device confirmed the break failure mode reported.The device was returned as two sections with the break 355mm from the strain relief.Based on the available information a definitive root cause cannot be determined.It was unknown if patient factors, handling or procedural techniques contributed to the reported event.Sleek pta rapid exchange (rx) dilatation catheter product family was reviewed the following sections are applicable: the sleek percutaneous transluminal angioplasty (pta) peripheral catheter family comprise a range of sizes of rapid exchange catheters for peripheral angioplasty.The catheter¿s proximal tubing is 304v stainless steel and the distal coaxial tubings are nylon co-polymer blend.The lumen of the shaft is used for the purpose of inflating and deflating the balloon.A second lumen at the tip is used for advancing the guidewire.Indications: the sleek catheters are intended for balloon dilatation of the femoral, popliteal and infra-popliteal arteries.These catheters are not designed to be used in the coronary arteries.Warnings: reuse, resterilization, reprocessing and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions: carefully inspect the catheter prior to use to verify that the catheter has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.Proper functioning of the catheter depends on its integrity.Care should be used when handling the catheter.Damage may result from kinking, stretching, or forceful wiping of the catheter.If the hypotube kinks prior to or during use the catheter should be discarded.No attempt should be made to straighten a kink in the hypotube.Storage: store in a cool, dark, dry place.Use the catheter prior to the ¿use by¿ date specified on the package.Directions for use: inspection and preparation remove the balloon sleeve by first withdrawing the shipping mandrel slightly and then slowly removing the sleeve while holding the catheter as close to the balloon as possible.If any resistance is felt, or if any stretching of the catheter is observed while removing the balloon sleeve, the product should not be used.The catheter should then be inspected for bends, kinks or stretched portions.Do not use if product damage is evident.Deflation and withdrawal simultaneously withdraw the dilatation catheter and guidewire from the guiding catheter/sheath.As the balloon exits the vessel, use a smooth, gentle, steady, motion.If resistance is felt upon removal then the balloon, guidewire and the guiding catheter/sheath should be removed together as a unit under fluoroscopic guidance, particularly if balloon rupture or leakage is known or suspected.This may be accomplished by firmly grasping the balloon catheter and guiding catheter/sheath as a unit and withdrawing both together, using a gentle twisting motion combined with traction.(b)(4).
 
Event or Problem Description
It was reported that the shaft of the pta balloon catheter allegedly broke during insertion through the sheath.The device was removed without incident.There was no reported difficulty in retracting the catheter through the introducer sheath.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLEEK RX PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI 
MDR Report Key7349597
Report Number9616666-2018-00029
Device Sequence Number555024
Product Code LIT
UDI-Device Identifier00801741132483
UDI-Public(01)00801741132483
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K072947
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2019
Device Model Number425-3022X
Device Catalogue Number425-3022X
Device Lot Number50137782
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2018
Initial Date Received by Manufacturer 02/20/2018
Supplement Date Received by Manufacturer05/15/2018
Initial Report FDA Received Date03/19/2018
Supplement Report FDA Received Date05/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-