| Catalog Number |
121820 |
| Medical Device Problem Codes |
Self-Activation or Keying (1557); Failure to Fire (2610); Device Misassembled During Manufacturing /Shipping (2912)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records and no medical images were provided to the manufacturer.The lot number of the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that during a kidney biopsy, the device allegedly failed to fire on the first sample pass.It was further reported that the procedure was completed with another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one monopathy biopsy needle was returned for evaluation.Functional testing was performed.The investigation was confirmed for self activation, as the cannula released on its own after locking into position.An x-ray noted the cannula spring out of position within the housing assembly.Upon disassembly, the cannula spring was noted to be wedged out of position within the two halves of the inner housing and preventing proper tang engagement.Therefore, the investigation was also confirmed for device disassembled.However, the investigation is inconclusive for failure to fire as a full functional evaluation could not be performed.The cannula spring out of position contributed to the identified self activation issue.It was possible that the cannula spring assembly issue contribute to the reported failure to fire issue.However, a definitive root cause could not be determined.Labeling review: a review of the product instructions for use (ifu) procedural instructions, indications, warnings, precautions, cautions, possible complications and contraindications showed that the product labeling is adequate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a kidney biopsy, the device allegedly failed to fire on the first sample pass.It was further reported that the procedure was completed with another device.There was no reported patient injury.
|
| |
|