• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH MNTR LS SCREW 3.5X36; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH MNTR LS SCREW 3.5X36; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION Back to Search Results
Model Number 188320336
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
I was covering a spine case at (b)(6) hospital ((b)(4)) with dr.(b)(6).We were performing a posterior spine fusion of c1-c2 using mountaineer.The case went well until the surgeon went to use the final tightener on the set screw (1883-41-200) while tightening, the set screw began to spin freely inside the pilyaxial head of the long shank screw 3.5mm x 36mm (1886-20-336).The surgeon attempted to use a separate set screw and realized the polyaxial head must have stripped.At this point we decided to back out the screw and replace it with the same item.Once this was completed a new set screw threaded perfectly and completed the construct.I saved the stripped screw and damaged set screws.Patient consequence? :no is the information being submitted for this complaint all the details that are known/available regarding this event? : yes.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Device evaluated by mfr: corrected.Visual examination of the returned screw features minimal signs of use beyond some light surface markings and discoloration.However, all of the threads inside of the tulip head have been torn off.This has left rounded edges where the threads were previously located.The damage is consistent across all of the threads of the tulip head.And only affects this one side.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.The root cause of the damage to the threads of the poly screw cannot be determined from the sample and the information provided.A potential root cause may be inadvertently cross threading an inner screw during insertion into the tulip head of this poly screw.This would place stress on the threads of the tulip head, potentially resulting in them becoming damaged or torn.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MNTR LS SCREW 3.5X36
Common Device Name
ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key7350116
Report Number1526439-2018-50247
Device Sequence Number11001000
Product Code NKG
UDI-Device Identifier10705034158167
UDI-Public(01)10705034158167
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K041203
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number188320336
Device Catalogue Number188320336
Device Lot Number5150W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Initial Date Received by Manufacturer 02/24/2018
Supplement Date Received by Manufacturer02/24/2018
04/20/2018
Initial Report FDA Received Date03/19/2018
Supplement Report FDA Received Date03/19/2018
04/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age88 YR
-
-