BARD ACCESS SYSTEMS DOT PROV 4F DL MAX TLS; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAY
|
Back to Search Results |
|
| Catalog Number |
S9274108D |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.A lot history review (lhr) of rebu0691 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (rebu0691) have been reported from two facilities.Device has not been returned, at this time.
|
| |
|
Event or Problem Description
|
|
The facility reported that the patient's line was placed on (b)(6) 2018.It was stated that the purple port completely slid off the line.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a catheter leak was confirmed and the failure is being further investigated under 3006260740-12-21-2017-001-c (z#-tbd).The product returned for evaluation was a 4fr d/l powerpicc provena catheter.The sampled contained white/gray dressing residue on the extension legs and molded joint.The purple luer hub had broken from the extension leg.Microscopic examination of the fracture surface showed that it was rough and granular and contained striation-like lines which are typical of fatigue based failures.Examination of the red luer hub, which remained attached to the opposite extension leg, found tearing of that extension leg.The tearing correlated to points where the extension tubing adhered to the luer hub.Tearing was not observed on regions of the extension tubing which were not adhered to the extension leg.The fracture features at the damage site appeared similar to fatigue damage and flexural fatigue may have been a contributing factor but the damage location and tacking of the extension tube to the luer wall suggest other, unknown, contributing factor(s) may have also been involved which will be further investigated under 3006260740-12-21-2017-001-c (z#-tbd).A lot history review (lhr) of rebu0691 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (rebu0691) have been reported from two facilities.
|
| |
|
Event or Problem Description
|
|
The facility reported that the patient's line was placed on (b)(6) 2018.It was stated that the purple port completely slid off the line.
|
| |
|
Search Alerts/Recalls
|
|
|