| Catalog Number |
0066814 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The immucor laboratory tested retention product on 05mar2018, which performed as expected.(b)(4).The immucor laboratory also tested a returned blood sample from the customer site on 05mar2018 against retention product, which yielded unexpected negative well outcomes.Therefore the customer reported event was reproduced.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a european (b)(6) customer site reported an unexpected negative antibody screen when using capture-r ready-screen (3) with a galileo neo instrument.
|
| |
|