WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.5MM THREADED GUIDE WIRE SPADE POINT 230MM; SCREW, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
900.723 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.Patient weight is not provided for reporting.(b)(4).Device malfunctioned intra-operatively and was not implanted / explanted complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that an initial surgery for implantation of a dynamic hip system (dhs) into the proximal femur was performed on (b)(6) 2018.During the surgery, the lag screw for the dhs plate was put in correctly, however, the plate was put in off-axis.Therefore, the plate had to be re-positioned, and as the surgeon was removing the connecting screw and guide shaft from the lag screw, the two instruments became bent (it was reported that the surgeon did not disconnect the instruments prior to taking them out) and the surgeon had to use pliers to take the instruments out.The same plate was put back into the patient without issue.A guide wire was also reported to be bent during the surgery.The patient was implanted with a lag screw, plate and three 4.5mm cortical screws (size 36, 42 and 48mm).There was no reported delay and the surgery was completed successfully.It was further reported that in sterile processing the connecting screw broke into two (2) pieces and the guide shaft, still in one piece, was bent trying to get them apart.This complaint is linked with (b)(4) for the post-operative event of broken/bent (connecting screw and guide shaft) instruments.Concomitant devices reported: dhs/dcs lag screw 12.7mm thread/90mm (part # 280.900, lot # unknown, quantity # 1), 4.5mm cortex screw (part # unknown, lot # unknown, quantity # 3).This complaint involves three (3) devices.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Date returned to manufacturer.Subject device has been received and is currently in the evaluation process.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
This is report 3 of 4 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A product development investigation was performed for the subject device: background: during the surgery, the dhs plate had to be re-positioned, and as the surgeon was removing the connecting screw and guide shaft from the lag screw, the two instruments became bent (it was reported that the surgeon did not disconnect the instruments prior to taking them out) and the surgeon had to use pliers to take the instruments out.The same plate was put back into the patient without issue.A guide wire was also reported to be bent during the surgery.Device condition: the 2.5mm threaded guide wire, part number 900.723, was received with a slight off axis bend.The origin of the bend is located approximately 140mm from the proximal end of the device (calipers: ca102p).The proximal end of the shaft also shows significant scraping such that the etching cannot be verified.The distal threads show wear consistent with insertion and removal.The balance of the device shows surface wear consistent with use and which would not impact the functionality.Thus, the complaint condition is confirmed and consistent with the reported condition.Replication of the complaint condition is not applicable as the device is already bent.Lot number review: a dhr review could not be performed as the lot number is unknown.Drawing review: the following drawings were reviewed.2.5mm guide wire.The outer diameter of the shaft on each side of the bend was inspected.The device diameter measured 2.47mm/2.49mm which is within the specification of 2.5mm +0/-0.1 per drawing (calipers: ca102p).During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.A material or material hardness check was not performed at cq as the lot number is unknown and the condition is consisted with an exceeding off-axis application of force; there is no indication that material or hardness contributed to this complaint.Conclusion: the complaint condition is confirmed as the wire was received with an off-axis bend.A device inspection and drawing review were performed as part of this investigation.There is no evidence of a manufacturing issue and no product design issues or discrepancies were observed that may have contributed to the complaint condition.No new malfunctions were observed during the course of this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|