Model Number ZMB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2012 and (b)(6) 2018.The lens remains implanted.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zmb00 18.5 diopter intraocular lens (iol) was implanted in the patient's right eye (od) on (b)(6) 2012.Post-operatively, the patient complained of not being able to read anymore and blurry vision.Reportedly, there is no plan to explant the lens, but the patient was prescribed eyeglasses.No additional information was provided.The patient had bilateral lenses implanted.This report will capture the lens implanted in the patient's right eye.A separate report will be filed for the left eye.
 
Additional Manufacturer Narrative
Correction: in follow up#2, date received by manufacturer was not correct as it should have indicated the following: april 16, 2018.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to the specifications.A search revealed that no similar complaints were received from this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: additional patient data stating that patient did have irregular corneas on both eyes (ou) pre-operatively, which is contraindicated for a tecnis multifocal (tmf) lens.In 2013 (1 year post-op), the patient did have 1+ posterior capsular opacification (pco) and 2+ pco on the right eye (od) and left eye (os) respectively, which likely contributed to the decreased visual acuity.No other post-op data was provided.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7351577
Report Number9614546-2018-00261
Device Sequence Number19106016
Product Code MFK
UDI-Device Identifier05050474535596
UDI-Public(01)05050474535596(17)140928
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/28/2014
Device Model NumberZMB00
Device Catalogue NumberZMB00U0185
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/20/2018
Supplement Date Received by Manufacturer03/12/2018
04/13/2018
05/15/2018
Initial Report FDA Received Date03/19/2018
Supplement Report FDA Received Date04/09/2018
05/12/2018
06/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-