WILLIAM COOK EUROPE COOK CELECT JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number IGTCFS-65-JUG-CELECT-PERM |
Device Problems
Fracture (1260); Difficult to Remove (1528); Obstruction of Flow (2423); Structural Problem (2506); Device Tipped Over (2589); Detachment of Device or Device Component (2907); Insufficient Information (3190)
|
Patient Problems
Edema (1820); Occlusion (1984); Internal Organ Perforation (1987); Thrombosis (2100); Perforation of Vessels (2135); Anxiety (2328); Depression (2361); Obstruction/Occlusion (2422); Device Embedded In Tissue or Plaque (3165); No Information (3190); Thrombosis/Thrombus (4440); Swelling/ Edema (4577)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Blank fields on this form indicate the information is unknown or unavailable.Catalog#: unknown but referred to as a cook celect filter.(b)(4).It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿celect filter implanted".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
It is alleged that "patient received a cook celect filter on (b)(6) 2009".It is alleged that patient was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
Exemption number e2016032e william cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: cook medical incorporated (cmi) (b)(4).Registration no.: (b)(4).Additional information: investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿celect - unable to retrieve, occlusive thrombus, venous insufficiency, edema, anxiety".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Ivc thrombotic occlusion as an outcome for cook ivc filters is addressed in the published scientific literature.Ivc thrombotic occlusion is defined as the presence of an occluding thrombus in the ivc after filter insertion and documented by ultrasound (us), ct, mr imaging or venography; this may be symptomatic or asymptomatic.Unknown if the reported anxiety is directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Manufacturer Narrative
|
Exemption number e2016032.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
|
Event Description
|
This additional information received on 06apr2018 as follows: pt allegedly received an implant on (b)(6) 2009 via the right internal jugular vein due to deep vein thrombosis.Pt alleges device is unable to be retrieved, edema due to clot burden, venous insufficiency, venous stasis, occlusive.Pt further alleges edema in both legs, anxiety.
|
|
Manufacturer Narrative
|
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: cook medical incorporated (cmi) (b)(4).Registration no.: (b)(4).Investigation: the following allegations have been investigated: fracture, venous insufficiency, edema.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Patient alleges fracture.
|
|
Manufacturer Narrative
|
Additional information: investigation: the following allegations have been investigated: vena cava (vc)/organ perforation, embedment, tilt, venous stasis, risk of future progressive filter failure, abuts duodenum, fear, stress, loss of enjoyment of life, depression.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported venous stasis, risk of future progressive filter failure, fear, stress, loss of enjoyment of life, and depression are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Patient further alleges tilt, and vena cava perforation.Additionally, patient alleges risk of future progressive filter failure, abuts duodenum, embedment, fear, stress, loss of enjoyment of life, depression.Report from computerized tomography (ct): "vasculature: there is an ivc filter in place in the appropriate location with approximately 7 degrees of leftward tilt of the apex.The posterior strut is fractured with the fractured portion embedded in the adjacent vertebral body.There is 4 mm of perforation of the tines along the anterior left and right aspect.The anterior strut abuts the duodenum without definite perforation.No evidence of ivc stenosis.The abdominal aorta is normal in caliber.".
|
|