| Catalog Number |
H965458950 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The end user hospital has been contacted in an effort to obtain more information about this event.The device sample is expected to be returned to angiodynamics for evaluation, but has not yet arrived.Upon completion of the investigation, a supplemental medwatch will be submitted.(b)(4).
|
| |
|
Event or Problem Description
|
|
As reported, an indwelling picc was removed and replaced due to a cracked hub.No patient injury or complication was reported.The used picc will be returned to angiodynamics for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history records was performed for the packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The recent angiodynamics complaint report was reviewed for the bioflo picc product family for the failure mode "hub broken/cracked." no adverse trends were indicated.No device sample was returned for evaluation, therefore an exact root cause for the reported issue is unable to be determined.A possible root cause for the reported crack is an over-tightened connection with a mating male luer (likely a needleless connector).Inadequate flushing of the device may also be a contributing factor.The directions for use (dfu) supplied with the reported picc device contain caution that "repeated overtightening may reduce hub connector life," and not to use hemostats to "secure or remove devices with luer lock hub connections." recommended flushing techniques are also stated in the dfu.(b)(4).
|
| |
|
Search Alerts/Recalls
|