LIVANOVA DEUTSCHLAND S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 48-50-00
Medical Device Problem Codes Thermal Decomposition of Device (1071); Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective ups module.The technician replaced the module to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The defective module was returned to livanova (b)(4) for further investigation.During the evaluation the reported issue could be reproduced.Two short circuits on the board could be identified as root cause of the reported issue.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event or Problem Description
Livanova (b)(4) received a report that a s5 system displayed an alarm and the customer reported burning smell after the alarm occurred during procedure.In addition the battery level indicator and percentage became blank there was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 SYSTEM
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7352385
Report Number9611109-2018-00237
Device Sequence Number1556381
Product Code DTQ
UDI-Device Identifier04033817900900
UDI-Public010403381790090011161121
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K071318
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48-50-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2018
Initial Report FDA Received Date03/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-