• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Insufficient Heating (1287)
Patient Problem No Patient Involvement (2645)
Event Date 02/21/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that the temperature of a heater-cooler system 3t is not stable and jumps sporadically by 2 degrees during priming.There was no patient involvement.
 
Manufacturer Narrative
A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and replaced all possible boards and ribbon cables, as well as the thermometer.However, the issue persisted.The components were installed back into the device and the unit was returned to livanova (b)(4) for investigation.The device was returned to the original equipment manufacturer for evaluation, which revealed that the connecting cable of the stepper motor was rubbing on the pump motor, causing a sporadic short circuit, which disrupted the electronics and caused the temperature display to become erratic.It could not be determined how the cable became loose.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7352400
MDR Text Key103182877
Report Number9611109-2018-00241
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Date Manufacturer Received03/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-