There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and replaced all possible boards and ribbon cables, as well as the thermometer.However, the issue persisted.The components were installed back into the device and the unit was returned to livanova (b)(4) for investigation.The device was returned to the original equipment manufacturer for evaluation, which revealed that the connecting cable of the stepper motor was rubbing on the pump motor, causing a sporadic short circuit, which disrupted the electronics and caused the temperature display to become erratic.It could not be determined how the cable became loose.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|