Model Number 48-50-00
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective e/p pack.The technician was not able to solve the issue in the field.Therefore the defective device was returned to livanova (b)(4) for further investigation and repair.During the investigation the reported failure could be reproduced and traced back to a defective hek board.As root cause faulty soldering joints could be identified.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event or Problem Description
Livanova (b)(4) received a report that a all display and control modules of a s5 system went blank during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 SYSTEM
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7352402
Report Number9611109-2018-00243
Device Sequence Number11001002
Product Code DTQ
UDI-Device Identifier04033817900900
UDI-Public010403381790090011161123
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K071318
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48-50-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2018
Initial Report FDA Received Date03/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-