| Catalog Number |
124114527 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
02/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Der states that the patient came to er with dislocated hip.Liner was removed and replaced.Doi: unknown; dor: (b)(6) 2018; right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: examination of the returned device confirms the reported event.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.
|
| |
|
Search Alerts/Recalls
|