• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - AIBONITO MICRO-VOLUME RADIATION STERILIZED EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - AIBONITO MICRO-VOLUME RADIATION STERILIZED EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2N3348
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the extension set tubing fell apart during set up.This event was further described as the end piece of the microbore tubing falling off.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed and it was noted that male luer lock adapter was missing.Dimensional inspection was performed with no issues noted.Further sample testing could not be performed as the sample was returned with a missing male luer lock adapter.The reported condition of the microbore tubing falling apart was verified.The cause of the reported condition was a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO-VOLUME RADIATION STERILIZED EXTENSION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - AIBONITO
aibonito
MDR Report Key7353052
Report Number1416980-2018-01458
Device Sequence Number1393391
Product Code FPA
UDI-Device Identifier00085412004822
UDI-Public(01)00085412004822
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K811078
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/22/2022
Device Catalogue Number2N3348
Device Lot NumberUR17F22044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2018
Initial Date Received by Manufacturer 02/23/2018
Supplement Date Received by Manufacturer04/02/2018
Initial Report FDA Received Date03/20/2018
Supplement Report FDA Received Date04/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-