MEDTRONIC IRELAND VALIANT CAPTIVIA; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number VAMF3232C150TU
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant stent graft system was implanted in a patient for the endovascular treatment of a 78mm dissection.It was reported that during the index procedure, the tip capture release handle would not turn to deploy the tip capture of the delivery system.It was noted that after multiple events, the proximal bare springs were released and the event was resolved.As per the physician, the cause of the event cannot be determined.No additional clinical sequelae were reported and the patient is fine.
 
Additional Manufacturer Narrative
Device evaluation summary: a kink was observed on the stent graft cover and middle member.The tip was slightly bent.No resistance was observed moving the tip release.There was no excessive glue evident in the portholes and no glue on the middle member.There was no evident of heath damage to the middle member/silicone.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7353515
Report Number2953200-2018-00397
Device Sequence Number16466676
Product Code MIH
UDI-Device Identifier00643169346987
UDI-Public00643169346987
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/23/2019
Device Model NumberVAMF3232C150TU
Device Catalogue NumberVAMF3232C150TU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2018
Initial Date Received by Manufacturer 02/22/2018
Supplement Date Received by Manufacturer06/27/2018
Initial Report FDA Received Date03/20/2018
Supplement Report FDA Received Date06/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-