| Catalog Number |
961-576 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information not provided due to (b)(6) patient privacy regulations.Device lot number, or serial number, unavailable.Device manufacturing date is dependent on lot number/serial number, therefore, unavailable.No parts have been received by the manufacturer for evaluation.Part not returned for analysis.
|
| |
|
Event or Problem Description
|
|
A manufacturer representative reported that, while in a spinal fusion, the clamp was found to be damaged.It was reported that a second clamp was used to complete the procedure.There was no reported delay to the procedure due to this issue.There was no reported impact on patient outcome.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot# 160505.Mfg date 2016-05-05.The spine clamp was returned for analysis; as reported, the adjustment screw had been broken off inside the clamp.If information is provided in the future, a supplemental report will be issued.
|
| |
|