MAUDE Adverse Event Report: NINEPOINT MEDICAL, INC. NVISIONVLE OPTICAL PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NINEPOINT MEDICAL, INC. NVISIONVLE OPTICAL PROBE Back to Search Results
Model Number 95301-M-20
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code Laceration(s) of Esophagus (2398)
Date of Event 02/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The 20 mm nvisionvle® optical probe was returned to ninepoint medical for evaluation.Balloon fold lines were observed, suggesting it was pulled about two-thirds of its length into the endoscope.The distal one-third of the balloon was not folded but inverted.This suggests that a pillow of air was present at the tip which prevented passage of the optical probe into the endoscope.The separation of the balloon from the sheath occurs when the balloon is insufficiently evacuated prior to removal.The evidence indicates that the majority of the balloon was inside the endoscope at the time of failure and thus no internal components of the optical probe had come into contact with the patient.The device history record was reviewed and no anomalies were identified that could have contributed to this event.Related warnings and precautions in the instructions for use specify: 1) when drawing the optical probe into the endoscope, if pillowing is observed, stop and either re-inflate and deflate, or remove the optical probe and scope as a system.2) use care when manipulating the endoscope while the balloon guide sheath is inflated.Extreme manipulations of the endoscope may result in breakage of the optical probe.3) this device will inflate to labeled diameter and therefore, should not be used in any anatomy where this size would be inappropriate.4) strictures, inflammatory disease or esophageal masses may prevent the adequate expansion of the optical probe.
 
Event or Problem Description
The physician selected a 20 mm nvisionvle® optical probe, which contains a 20 mm balloon, for the oct imaging evaluation.The optical probe was inserted through the working channel of the endoscope, inflated, and the imaging procedure was performed successfully.At the end of the imaging procedure, the technician deflated the 20 mm balloon and began to remove the optical probe.At this point, it was noted that the optical probe had difficulty being removed from the endoscope.The optical probe was removed from endoscope, however they noticed that the balloon was detached from the optical probe and lodged in the endoscope channel.The endoscope with lodged balloon was removed from patient.It was observed that the optical probe was lodged in the endoscope and the proximal end of the balloon was still partially inflated.It is presumed that the balloon was not fully deflated before trying to remove the optical probe.The optical probe's balloon was removed from the endoscope channel and the endoscope was reinserted.Upon further examination of patient after the oct imaging procedure, two bleeding mucosal tears near the esophagogastric junction were noticed.The bleeding was stopped with 2 hemostatic clips at the site.There were no strictures or hernia noticed in that area.The physician was not concerned about the bleeding, and the procedure including subsequent biopsy was completed with no complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NVISIONVLE OPTICAL PROBE
Common Device Name
NVISIONVLE OPTICAL PROBE
Manufacturer (Section D)
NINEPOINT MEDICAL, INC.
12 oak park drive
bedford MA 01730
MDR Report Key7354822
Report Number3008805841-2018-90002
Device Sequence Number13775971
Product Code NQQ
UDI-Device Identifier00859591006072
UDI-Public(01)00859591006072(17)180930(10)005263
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K153479
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/30/2018
Device Model Number95301-M-20
Device Catalogue Number95301-M-20
Device Lot Number5263
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2018
Initial Report FDA Received Date03/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-