• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMERIEUX INC. VITEK® 2 NH TEST KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMERIEUX INC. VITEK® 2 NH TEST KIT Back to Search Results
Catalog Number 21346
Medical Device Problem Code Incorrect Or Inadequate Test Results (2456)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
A customer in the united states reported a misidentification of (b)(6) as neisseria meningitides in association with the vitek® 2 nh test kit.The customer stated the initial identification of n.Meningitides was reported to the physician.The physician requested that the isolate be retested as they believed it to be (b)(6), and the patient was being treated for it.The customer performed repeat testing with the nh card from the original plate, as there was not enough growth for the required mcfarland range.A low discrimination result was received, which did not include (b)(6).The isolate was sent out to a reference lab which reported that the isolate was dead and no identification could be made.The customer stated the isolate is no longer available.There is no indication or report from the hospital or treating physician to biomérieux that the discrepant result led to any adverse event related to the patient's state of health.A biomérieux internal investigation will be initiated.
 
Additional Manufacturer Narrative
A customer in the united states reported a misidentification of neisseria gonorrhea as neisseria meningitides in association with the vitek® 2 nh test kit.An internal biomérieux investigation was performed.The customer reported setting up an 18-24 hour old isolate from chocolate agar that had been incubated at 36c in co2.Three (3) lab reports were submitted.The lab report labeled initial run showed an excellent identification of n.Meningitidis with five (5) atypical reactions (arga-, ggt+, proa-, tyra-, appa-) for an identification of n.Gonorrhoeae according the nh knowledge base.A second lab report was noted to be a re-test of the isolate.A note appeared on the lab report that the test was "run on an old plate - did not have enough organism to get required mcfarland".The lab report showed a low discrimination result for n.Cinerea and n.Meningitidis with five (5) atypical negative reactions (arga, proa, tyra, appa, dglu) for an identification of n.Gonorrhoeae according the nh knowledge base.A third lab report was noted to be a known isolate and not the patient isolate.The lab report showed a low discrimination result for n.Cinerea, n.Elongata, and n.Gonorrhoeae, which would be considered a correct result for n.Gonorrhoeae.An increased number of atypical reactions can indicate contamination, mixed culture, compromised viability of the strain, use of non-recommended media or other user set up errors, or an atypical strain.For the repeat testing of the isolate, the potential misidentification was most likely due to the age of culture being out of range and insufficient mcfarland density.However, without the strain or raw data it is not possible to further evaluate the cause of the misidentification.For fastidious species such as n.Gonorrhoaeae, it is important to limit the amount of time the culture is out of co2.This species will begin to show decreased viability and therefore, less reactivity in the nh card when out of co2 for an extended period of time.A note will be added in 8.01 software that will state "possibility of n.Gonorrohoeae" when an identification of n.Cinerea is reported.Vitek 2 nh test kit lot #2450361203 met final qc release criteria and passed qc performance testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITEK® 2 NH TEST KIT
Common Device Name
VITEK® 2 NH TEST KIT
Manufacturer (Section D)
BIOMERIEUX INC.
595 anglum road
st. louis MO 63042
MDR Report Key7355105
Report Number1950204-2018-00111
Device Sequence Number13746347
Product Code JST
UDI-Device Identifier03573026144357
UDI-Public03573026144357
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
C1 EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/22/2018
Device Catalogue Number21346
Device Lot Number2450361203
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2018
Supplement Date Received by Manufacturer05/01/2018
Initial Report FDA Received Date03/20/2018
Supplement Report FDA Received Date05/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-