• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; TURON PRIMARY HUMERAL STEM, SZ 16

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; TURON PRIMARY HUMERAL STEM, SZ 16 Back to Search Results
Model Number 520-01-016
Health Effect - Clinical Code Excessive Tear Production (2235)
Date of Event 03/13/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to subscapularis failure.The surgeon converted to a reverse shoulder.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to a subscapularis failure.The previous surgery and the revision detailed in this investigation occurred over 1 year and 11 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to subscapularis failure.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: excessive range-of-motion, soft tissue impingement, prolonged usage of overhead activities, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PRIMARY HUMERAL STEM, SZ 16
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7355326
Report Number1644408-2018-00241
Device Sequence Number13958970
Product Code KWS
UDI-Device Identifier00888912024921
UDI-Public(01)00888912024921
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model Number520-01-016
Device Catalogue Number520-01-016
Device Lot Number458G1047
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/13/2018
Supplement Date Received by Manufacturer03/28/2018
Initial Report FDA Received Date03/20/2018
Supplement Report FDA Received Date03/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000, LOT 878C1568; 520-01-250, LOT 193G1107; 520-50-022, LOT 931C1067; 520-00-000, LOT 878C1568; 520-01-250, LOT 193G1107; 520-50-022, LOT 931C1067
Outcome Attributed to Adverse Event Required Intervention;
-
-