U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - SINGAPORE INTERLINK SYSTEM; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - SINGAPORE INTERLINK SYSTEM; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C7451
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cause of the reported condition was the user error attempting to use an interlink system secondary medication set (2c7451) and solution set with duo-vent spike (act5435) together.The two devices are not compatible with each other as the act5435 has a needle based y-site and the 2c7451 uses a male luer at the end of the set.Proper user instructions are addressed in the labeling for an interlink system secondary medication set, which is shipped with the device.The label warns the user that the secondary medication set interlink system is for use with interlink system continu-flo solution sets.The label also includes instruction for use which state to insert the cannula into an interlink y-injection site on continu-flo set.Proper user instructions are addressed in the labeling for a solution set with duo-vent spike (act5435), which is shipped with the device.The label includes instruction for use which state to access the injection site with siliconized 18 gauge (or smaller bore) needle.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an interlink secondary medication set would not connect to the access port of a solution set.During setup, the customer was unable to puncture the port of the primary solution set with the secondary set.This event occurred during patient infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK SYSTEM
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
singapore
MDR Report Key7355642
Report Number1416980-2018-01482
Device Sequence Number1220819
Product Code FPA
UDI-Device Identifier00085412002576
UDI-Public(01)00085412002576
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123868
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2022
Device Catalogue Number2C7451
Device Lot NumberSR17F06037
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2018
Initial Report FDA Received Date03/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-