Additional manufacturer narrative: (b)(4) is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number (b)(4).This report is being submitted due to a retrospective review conducted under capa (b)(4).Device by manufacturer: on (b)(6) 2015, an fse was dispatched to the customer facility to address the reported event.Fse began to troubleshoot the issue over the phone.Fse instructed customer to unplug the power var unit and not use it.Customer informed fse that they have another power var unit from the g7 unit and will use that one.Customer will contact power var manufacturer to see if it can be repaired or replaced.The customer ran quality control (qc) in which all results were within acceptable range and verified that the analyzer was functioning correctly.The g8 analyzer was functioning as intended and the errors cleared.No further action required by field service.The probable cause of the event was power var making noise and seems to be overheating.
|
On (b)(6) 2015, a customer reported power var is making alot of noise and seems to run hot with their g8 analyzer.The customer is unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|