BARD PERIPHERAL VASCULAR, INC. SEEKER; SUPPORT CATHETER Back to Search Results
Model Number SK15035M
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during treatment of a heavily calcified lesion in the right sfa the tip of the support catheter allegedly detached and migrated to the profunda femoris artery.The health care provider (hcp) attempted to snare the detached tip but was unsuccessful.It was further reported that the hcp decided to leave the detached tip in situ as there was no potential health hazard to the patient.The procedure was concluded, and no further treatment was provided.There was no reported patient injury.
 
Event or Problem Description
It was reported that during treatment of a heavily calcified lesion in the right sfa the tip of the support catheter allegedly detached and migrated to the profunda femoris artery.The health care provider (hcp) attempted to snare the detached tip but was unsuccessful.It was further reported that the hcp decided to leave the detached tip in situ as there was no potential health hazard to the patient.The procedure was concluded, and no further treatment was provided.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned.The investigation is confirmed for tip detachment as the 1.8cm of the catheter distal end is detached and not returned.Per the reported event the lesion was heavily calcified and the catheter was not being used for a recanalization procedure, this may have contributed to the event.However, the root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: while retracting the catheter, if resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or catheter separation.Precautions: if resistance is felt during post procedure withdrawal of the catheter, it is recommended to remove the catheter, introducer sheath and guidewire (as necessary) as a single unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEEKER
Common Device Name
SUPPORT CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7356162
Report Number2020394-2018-00261
Device Sequence Number16533120
Product Code DQY
UDI-Device Identifier00801741067921
UDI-Public(01)00801741067921
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K103367
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/12/2020
Device Model NumberSK15035M
Device Catalogue NumberSK15035M
Device Lot NumberVTCN0434
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Initial Date Received by Manufacturer 02/21/2018
Supplement Date Received by Manufacturer03/27/2018
Initial Report FDA Received Date03/20/2018
Supplement Report FDA Received Date04/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
-
-