The scope was returned to olympus for evaluation.
The evaluation found the distal end cover insulation in critical condition.
The distal end cover was found cracked and burned.
The bending section rubber glue was found cracked.
In addition, the insertion tube was found with minor scratches, buckles, and dents.
The light guide tube and the video cable were also found with minor scratches.
The scope was serviced and returned to the user facility.
Based on the investigation findings, the exact cause of the reported event could not be conclusively determined; however, improper maintenance of the scope could not be ruled out as a contributing factor to the reported event.
The instruction manual provides warning and caution statements in an effort to prevent equipment damage and patient harm.
¿the probability of failure of the endoscope and ancillary equipment increases as the number of procedures performed and/or the total operating hours increase.
In addition to the inspection before each procedure, the person in charge of medical equipment maintenance in each facility should inspect the items specified in this manual periodically.
Before each case, prepare and inspect this instrument as instructed below.
Inspect other equipment to be used with this instrument as instructed in their respective instruction manuals.
If this instrument malfunctions, do not use it.
Return it to olympus for repair.
Anytime you observe an irregularity in an endoscope function, stop the examination immediately and slowly remove the endoscope while viewing the endoscopic image if possible.
Using an endoscope that is not functioning properly may compromise patient or operator safety and may result in more severe equipment damage.
¿.
|